PoS - Proceedings of Science
Volume 322 - 5th International Conference on Micro-Pattern Gas Detectors (MPGD2017) - Main session
The micro-RWELL detector
M. Poli Lener*, G. Bencivenni, G. Felici, M. Gatta, G. Morello and R. De Oliveira
Full text: pdf
Pre-published on: August 06, 2019
Published on: August 21, 2019
Abstract
The micro-RWELL is a compact spark-protected single amplification stage Micro-Pattern-Gaseous-Detector (MPGD). The detector amplification stage is realized with a polyimide structure, micropatterned with a dense matrix of blind-holes, integrated into the readout structure. The anode is formed by a thin Diamond Like Carbon (DLC) resistive layer separated by an insulating glue layer from the readout strips. The introduction of the resistive layer strongly suppressing the transition from streamer to spark gives the possibility to achieve large gains (> 104), without significantly affecting the capability to be efficiently operated in high particle fluxes.
In this work we present the results of a systematic study of the micro-RWELL performance as a function of the DLC resistivity. The tests have been performed either with collimated 5.9 keV X-rays or with pion and muon beams at the SPS Secondary Beamline H4 and H8 at CERN.
DOI: https://doi.org/10.22323/1.322.0019
How to cite

Metadata are provided both in "article" format (very similar to INSPIRE) as this helps creating very compact bibliographies which can be beneficial to authors and readers, and in "proceeding" format which is more detailed and complete.

Open Access
Creative Commons LicenseCopyright owned by the author(s) under the term of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.