Main Image
Volume 343 - Topical Workshop on Electronics for Particle Physics (TWEPP2018) - Asic
Characterization of GEMINI, a 16-channels programmable readout interface for Triple-GEM detectors in 180nm CMOS
L. Mangiagalli,* G. Croci, M. De Matteis, D. Tagnani, G. Corradi, F. Murtas, G. Gorini, A. Baschirotto
*corresponding author
Full text: pdf
Pre-published on: 2019 May 21
Published on: 2019 July 25
Abstract
The recent advances in GEM detector development has led to the requirement of a custom readout to fully exploit the advantages of this technology. GEM detectors can be realized with various shapes, also irregular, and high number of channels. GEMINI has been specifically designed to work with Triple-GEM detectors and it integrates 16 channels to perform readout with both analog and digital signal with Time over Threshold. GEMINI also allows to set a different threshold for every channel to compensate channel inhomogeneity. This work compares simulations with lab measurements and presents results of the imaging of an X-ray source performed with GEMINI.
DOI: https://doi.org/10.22323/1.343.0165
Open Access
Creative Commons LicenseCopyright owned by the author(s) under the term of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.