Calibration and Performance of the CMS Electromagnetic Calorimeter in LHC Run 2
D. Konstantinov* on behalf of the CMS Collaboration
Pre-published on:
February 02, 2021
Published on:
April 15, 2021
Abstract
Many physics analyses using the Compact Muon Solenoid (CMS) detector at the LHC require accurate, high-resolution electron and photon energy measurements. The excellent energy resolution is crucial for studies of Higgs boson decays with electromagnetic particles in the final state, as well as searches for very high mass resonances decaying to energetic photons or electrons. The CMS electromagnetic calorimeter (ECAL) is a fundamental instrument for these analyses and its energy resolution is crucial for the Higgs boson mass measurement. Recently the energy response of the calorimeter has been precisely calibrated exploiting the full Run 2 data, aiming at a legacy reprocessing of the data. A dedicated calibration of each detector channel has been performed with physics events exploiting electrons from W and Z boson decay, photons from pi0 decay, and from the azimuthally symmetric energy distribution of minimum bias events. The calibration strategies that have been implemented and the excellent performance achieved by the CMS ECAL with the ultimate calibration of Run 2 data, in terms of energy scale stability andenergy resolution, are presented.
DOI: https://doi.org/10.22323/1.390.0804