PoS - Proceedings of Science
Volume 379 - The 18th International Workshop on Polarized Sources, Targets, and Polarimetry (PSTP2019) - Solid Polarized Targets
Magnetic Field Requirements for the CLAS12 Polarized Target
V. Lagerquist
Full text: pdf
Published on: September 23, 2020
Abstract
Upcoming spin structure experiments in Hall B at Jefferson Lab will employ a new dynamically polarized target inside the CLAS12 detector system. Protons and deuterons in irradiated NH$_3$ and ND$_3$ will be polarized at 1 K using the 5 T field of the CLAS12 solenoidal magnet. For optimum polarization, the field uniformity requirements are around 100 ppm over the volume of the target sample. I will discuss methods to improve field uniformity utilizing thin superconducting shim coils integrated within the 1 K refrigerator. I will also demonstrate that this method to adjust the 5 T field also enables the simultaneous opposite polarization of two adjacent target cells.
DOI: https://doi.org/10.22323/1.379.0055
How to cite

Metadata are provided both in "article" format (very similar to INSPIRE) as this helps creating very compact bibliographies which can be beneficial to authors and readers, and in "proceeding" format which is more detailed and complete.

Open Access
Creative Commons LicenseCopyright owned by the author(s) under the term of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.