PoS - Proceedings of Science
Volume 390 - 40th International Conference on High Energy physics (ICHEP2020) - Posters: Neutrino Physics
Movable Thermometer System in ProtoDUNE
R. Dharmapalan* and  On behalf of the DUNE Collaboration
Full text: pdf
Pre-published on: January 04, 2021
Published on: April 15, 2021
Abstract
The movable temperature profiler is a 7 m vertical array of 24 sensors that measures cryogenic temperatures with a precision of a few mK. This precision is necessary to monitor the efficiency of re-circulation and purification of liquid-argon inside large liquid-argon based neutrino detectors. Liquid argon temperature impacts electron (signal) drift velocity, flow, purity distribution and thus the overall energy calibration. The temperature profiler is motorized and moves vertically, while in the detector, and cross-calibrates neighboring sensors. The temperature offsets between each sensor cancel the effects of electromagnetic noise. This poster reports on the temperature measurements and such in-situ cross-calibrations at ProtoDUNE (single phase) at CERN.}
DOI: https://doi.org/10.22323/1.390.0197
How to cite

Metadata are provided both in "article" format (very similar to INSPIRE) as this helps creating very compact bibliographies which can be beneficial to authors and readers, and in "proceeding" format which is more detailed and complete.

Open Access